Technical data

1 Installation
12 Agilent Multimode for 6210/6220 TOF Installation and Maintenance Guide
Figure 2 Carefully remove the chip from the socket
5 In the Firmware Upgrade Kit, find the chip with the part number
G1978-80067.
6 Find the correct orientation of the chip so that it fits into the socket. Press
the chip into the socket.
7 Locate the Power Distribution Board (PDB) that was removed in “Step 2.
Remove the TOF E-Tub Electronics” on page 10. Place the PDB on a static
free surface.
8 Using the same technique that was used to remove the chip on the Analyzer
Board, remove the chips from the U6 and U18 sockets. See Figure 3.
CAUTION
Be sure that the chip is completely in the socket. Otherwise, when the board is
energized the chip will be destroyed.