Technical data
5 Indexing and Sample Processing for Multiplexed Sequencing
Step 2. Purify the amplified captured libraries using AMPure XP beads
78 SureSelect
XT
Target Enrichment System for Illumina Multiplexed Sequencing
15 Seal the sample wells, then mix well on a vortex mixer and briefly spin
to collect the liquid.
16 Incubate for 2 minutes at room temperature.
17 Put the plate or strip tube in the magnetic stand and leave for
2 minutes or until the solution is clear.
18 Remove the cleared supernatant (approximately 30 µL) to a fresh well.
You can discard the beads at this time.
Stopping Point If you do not continue to the next step, store the libraries at 4°C for up
to one week or at –20°C for longer periods.