Technical data
Sample Preparation (200 ng DNA Samples) 3
Step 9. Amplify the adaptor-ligated library
SureSelect
XT
Target Enrichment System for Illumina Multiplexed Sequencing 53
Step 9. Amplify the adaptor-ligated library
This step uses the components listed in Table 27. Thaw the reagents listed
below and keep on ice.
Table 27 Reagents for pre-capture PCR amplification
Component Storage Location
SureSelect Primer (brown cap) SureSelect XT Library Prep Kit ILM, –20°C
SureSelect ILM Indexing Pre-Capture PCR
Reverse Primer (clear cap)
SureSelect Target Enrichment Kit ILM Indexing
Hyb Module Box 2, –20°C
Herculase II Fusion DNA Polymerase (red cap) Herculase II Fusion DNA Polymerase kit, –20°C
5× Herculase II Reaction Buffer (clear cap) Herculase II Fusion DNA Polymerase kit
*
, –20°C
* Do not use the PCR Reaction Buffer or dNTP mix from any other kit.
100 mM dNTP Mix (green cap) Herculase II Fusion DNA Polymerase kit*, –20°C