Specifications
5-6
5.1.3 Causes of negative D problem
When measuring the dissipation factor (D) of a low loss capacitor, the impedance measuring instru-
ment may sometimes display a negative D value despite the fact that the real dissipation factor must
be a positive value. A negative D measurement value arises from a measurement error for a small
resistance component of the measured impedance. In this section, we discuss the causes of negative
D and the methods for minimizing the measurement errors that lead to the negative D problem. Five
typical causes of negative D problem are:
• Instrument inaccuracy
• Contact resistance in the 4TP or 5T configuration
• Improper short compensation
• Improper cable length correction
• Complicated residuals
Note: The following discussion also applies to a negative Q problem because the Q factor is the
reciprocal of D.
D measurement error due to instrument inaccuracy
If a DUT has a low D value compared with the D measurement accuracy (allowable D measurement
error) of the instrument, a measured dissipation factor may become a negative value. Figure 5-8
shows how the D measurement accuracy of instrument impacts a negative D value. For example,
when D measurement accuracy (of instrument A) is ±0.001 for a low-loss capacitor that has a dissi-
pation factor of 0.0008, the impedance measurement error is represented by a dotted circle on the
vector plane as shown in Figure 5-8. The shaded area of the dotted circle exists on the left side of
reactance axis (X axis.) This shaded area represents the negative D area in which the resistance
component of the measured impedance is a negative value. The allowable D value range is from
–0.0002 to 0.0018. In this case, there is possibility that a negative D value is displayed. If the D mea-
surement accuracy (of instrument B) is ±0.0005, the measured impedance vector is within the solid
circle as shown in Figure 5-8. The negative D value is not displayed because the allowable D value
range is from 0.0003 to 0.0013. Accordingly, an impedance measuring instrument with the best pos-
sible accuracy is required for avoiding negative D display in low dissipation factor measurements.
Figure 5-8. Negative D measurement value due to measurement inaccuracy
Example:
D = 0.0008
(at specific measurement conditions)
Instrument D accuracy Possible readout
A
±
0.001 –0.0002 to 0.0018
B
±
0.0005 0.0003 to 0.0013
R
Negative D
X
0.001 accuracy
0.0005 accuracy
Impedance
vector