Specifications

3.2.2 User-fabricated test fixtures
If the DUT is not applicable to Agilent-supplied test fixtures, create an application-specific test
fixture. Key points to consider when fabricating a test fixture are:
(1) Residuals must be minimized. To minimize the residuals, the 4TP configuration should be main-
tained as close as possible to the DUT. Also, proper guarding techniques will eliminate the
effects of stray capacitance. For details, refer to “Practical guarding techniques” in Section 3.4.
(2) Contact resistance must be minimized. Contact resistance will cause additional error. In the case of
the 2T configuration, it directly affects the measurement result. The contact electrodes should
hold the DUT firmly and should always be clean. Use a corrosion-free material for the elec-
trodes.
(3) Contacts must be able to be opened and shorted. Open/short compensation can easily reduce the
effects of the test fixture's residuals. To perform an open measurement, the contact electrodes
should be located the same distance apart as when the DUT is connected. For the short mea-
surement, a lossless (low impedance) conductor should be connected between the electrodes, or
the contact electrodes should be directly interconnected. If the 4T configuration is kept to the
electrodes, make the connections of current and potential terminals, and then make an open or
short as shown in Figure 3-9.
Figure 3-9. User-fabricated test fixture open/short methods
Hc
Hp
Lc
Lp
DUT electrodes
DU T
Contact
electrodes
Hc
Hp
Lc
Lp
(a) DUT connection
(b) OPEN measurement
OR
OR
(c) SHORT measurement
Hc
Hp
Lc
Lp
Low loss
conductor
Low loss
conductors
Hc
Hp
Lc
Lp
Hc
Hp
Lc
Lp
3-8