Specifications
3.0 Fixturing and Cabling
Connecting a DUT to the measurement terminals of the auto-balancing bridge instrument requires a
test fixture or test cables. The selection of the appropriate test fixtures and cables, as well as the
techniques for obtaining the optimum DUT connection configuration, are important for maximizing
the total measurement accuracy. This section introduces the basic theory and use of each connec-
tion configuration, focusing on the auto-balancing bridge instrument. In RF impedance measure-
ments, the usable connection configuration is the two-terminal (2T) configuration only. Since the
measurement technique for RF impedance is different from that for LF, it is described separately
after the discussion of the auto-balancing bridge instrument.
3.1 Terminal configuration
An auto-balancing bridge instrument is generally equipped with four BNC connectors, Hcur, Hpot,
Lpot, and Lcur, as measurement terminals (see Figure 3-1.) These terminals are conventionally
named "UNKNOWN" terminals. There are several connection configurations used to interconnect a
DUT with the UNKNOWN terminals. Because each method has advantages and disadvantages, the
most suitable method should be selected based on the DUT’s impedance and required measurement
accuracy.
Figure 3-1. Measurement terminals of auto balancing bridge instrument
L
cur
: Low current
L
pot
: Low potential
H
po t
: High potential
H
cur
: High current
Precision LCR Meter
3-1