Preventive Maintenance
Instrument Installation
To install or re-install an instrument in a system, use one of the following
procedures.
Required tools
#2 Phillips screwdriver
#2 POZIDRIV screwdriver
system installation guide
Standard rack instrument
To install an instrument
Step Notes
1 Slide the instrument gently into the rack.
2 Insert the screws in the rack ears.
Most cables are fairly easy to remove and
reconnect and have no special
considerations (besides making sure you
put the cables back in the right place).
Semirigid cables require more care,
especially when reconnecting them.
Make sure all semirigid cables, on the
front and back of an instrument are fully
disconnected before removing the unit.
3 To reconnect the semirigid cables, carefully
align them before you insert the male
connector.
Do not insert the male pin in at an angle
or you will damage the female connector.
RF connector center pins are very
delicate, and if damaged must be
replaced. System performance may be
greatly impaired if there is a bad RF
connector.
4 Turn on system power and restart the system
computer if necessary.