User`s guide

Table Of Contents
Introduction and Measurement
2
Agilent E5505A User’s Guide 43
Congratulations
You have completed a phase noise measurement. This measurement of the test
set’s low noise amplifier provides a convenient way to verify that the system
hardware and software are properly configured for making noise
measurements. If your graph looks like that in Figure 11, you can be confident
that your system is operating normally.
Learning more
Continue with this demonstration by turning to Chapter 4, “Expanding Your
Measurement Experience to” learn more about performing phase noise
measurements.
Table 3 Parameter data for the N5500A confidence test example
Step Parameters Data
1 Type and Range Tab
Measurement Type
Start Frequency
Stop Frequency
Minimum Number of Averages
FFT Quality
Swept Quality
Baseband Noise (using a test set)
10 Hz
100 E + 6 Hz
1
4
Fast
Fast
2 Cal Tab
Gain preceding noise input 0 dB
3 Block Diagram Tab
Noise Source Test Set Noise Input
4 Test Set Tab
Input Attenuation
LNA Low Pass Filter
LNA Gain
DC Block
PLL Integrator Attenuation
0 dB
20 MHz (Auto checked)
Auto Gain (Minimum Auto Gain –14 dB)
Not checked
0 dBm