User`s guide

Table Of Contents
Baseband Noise Measurement Examples
13
Agilent E5505A User’s Guide 323
Table 50 Parameter data for the baseband without using a test set measurement
Step Parameters Data
1 Type and Range Tab
Measurement Type
Start Frequency
Stop Frequency
Averages
Quality
Baseband Noise (without using a test set)
10 Hz
100 E + 6 Hz
4
Normal
2Cal Tab
Gain preceding noise input
0 dB
3 Block Diagram Tab
Noise Source
Test Set Noise Input
5Graph Tab
Title
Graph Type
X Scale Minimum
X Scale Maximum
Y Scale Minimum
Y Scale Maximum
Normalize trace data to a:
Scale trace data to a new carrier
frequency of:
Shift trace data DOWN by:
Trace Smoothing Amount
Power present at input of DUT
Baseband Noise without using a Test Set
Baseband (dBV)
10 Hz
100 E + 6 Hz
0 dBc/Hz
–200 dBV/Hz
1 Hz bandwidth
1 times the current carrier frequency
0 dB
0
0 dB