User`s guide

Table Of Contents
AM Noise Measurement Examples
12
Agilent E5505A User’s Guide 313
Table 48 Parameter data for the AM noise using an N5500A Option 001
Step Parameters Data
1 Type and Range Tab
Measurement Type
Start Frequency
Stop Frequency
Averages
FFT Quality
Swept Quality
AM Noise
10 Hz
100 E + 6 Hz
4
Fast
Fast
2Sources Tab
Carrier Source Frequency
Carrier Source Power
Carrier Source Output is
connected to:
Detector Input Frequency
600 E + 6 Hz
20 dBm
Test Set
600 E +6 Hz
3Cal Tab
Detector Constant
Known Spur Parameters
Offset Frequency
Amplitude
Use internal automatic self-calibration
1 Hz
-130 dBc
4 Block Diagram Tab
Source
AM Detector
Down Converter
Manual
Test Set AM Detector
None
5Test Set Tab
Input Attenuation
LNA Low Pass Filter
LNA Gain
Detector Maximum Input Levels
Microwave Phase Detector
RF Phase Detector
AM Detector
Ignore out-of-lock conditions
Pulsed Carrier
DC Block
Analyzer View
PLL Integrator Attenuation
Auto checked
Auto checked
Auto Gain
0 dBm
0 dBm
0 dBm
Not checked
Not checked
Not checked
Baseband
0.00 dBm
6Downconverter Tab Does not apply to this measurement example.