282 Agilent E5505A User’s Guide
10
FM Discriminator Measurement Examples
Table 45 Parameter data for the rate and deviation calibration example
Step Parameters Data
1 Type and Range Tab
Measurement Type
Start Frequency
Stop Frequency
Minimum Number of Averages
FFT Quality
Swept Quality
Absolute Phase Noise (using an FM Discriminator)
10 Hz
100 E + 6 Hz
4
Normal
Fast
2Sources Tab
Carrier Source
Frequency
Power
Carrier Source is Connected to:
Detector Input
Frequency
1.027 E + 9 Hz
19 dBm
Te s t S e t
1.027 E + 9 Hz
3Cal Tab
FM Discriminator Constant
Current Phase Detector
Constant
Know Spur Parameters
Offset Frequency
Amplitude
Calibration Source
Frequency
Power
Frequency Modulation
FM Rate
FM Deviation
Derive Constant from FM rate and deviation
82.25 E-9
1 E3
-6 dBc
1.027 E + 9 Hz
16 dBm
20 E +3 Hz
10 E +3 Hz
4 Block Diagram Tab
Carrier Source
Phase Shifter
DUT in Path
Phase Detector
Adjust the Quadrature by
adjusting the
Manual
Manual
checked
Automatic Detector Selection
phase shifter
5 Test Set Tab The test set parameters do not apply to this
measurement example.
6 Downconverter Tab The downconverter parameters do not apply to this