8
Residual Measurement Examples
4 Block Diagram Tab
Carrier Source
Phase Shifter
DUT in Path
Phase Detector
Adjust the Quadrature by
adjusting the
Manual
Manual
checked
Automatic Detector Selection
phase shifter
5Test Set Tab
Input Attenuation
LNA Low Pass Filter
LNA Gain
DC Block
PLL Integrator Attenuation
0 dB
20 MHz (Auto checked)
Auto Gain (Minimum Auto Gain –14 dB)
Not checked
0 dBm
6
Dowconverter Tab
The downconverter parameters do not apply to this
measurement example.
7Graph Tab
Title
Graph Type
X Scale Minimum
X Scale Maximum
Y Scale Minimum
Y Scale Maximum
Normalize trace data to a:
Scale trace data to a new
carrier frequency of:
Shift trace data DOWN by:
Trace Smoothing Amount
Power present at input of DUT
E5505A Residual Phase Noise Measurement
@ 1 GHz.
Single-sideband Noise (dBc/Hz)
10 Hz
100 E + 6 Hz
0 dBc/Hz
- 180 dBc/Hz
1 Hz bandwidth
1 times the current carrier frequency
0 dB
0
0 dB
Table 38 Parameter data for the amplifier measurement example (continued)
Step Parameters Data