User`s guide
Table Of Contents
- Title Page
- Contents
- Getting Started
- Introduction and Measurement
- Phase Noise Basics
- Expanding Your Measurement Experience
- Starting the Measurement Software
- Using the Asset Manager
- Using the Server Hardware Connections to Specify the Source
- Setting GPIB Addresses
- Testing the 8663A Internal/External 10 MHz
- Testing the 8644B Internal/External 10 MHz
- Viewing Markers
- Omitting Spurs
- Displaying the Parameter Summary
- Exporting Measurement Results
- Absolute Measurement Fundamentals
- Absolute Measurement Examples
- Residual Measurement Fundamentals
- What is Residual Noise?
- Assumptions about Residual Phase Noise Measurements
- Calibrating the Measurement
- Measurement Difficulties
- Residual Measurement Examples
- FM Discriminator Fundamentals
- FM Discriminator Measurement Examples
- AM Noise Measurement Fundamentals
- AM Noise Measurement Examples
- Baseband Noise Measurement Examples
- Evaluating Your Measurement Results
- Advanced Software Features
- Reference Graphs and Tables
- Approximate System Noise Floor vs. R Port Signal Level
- Phase Noise Floor and Region of Validity
- Phase Noise Level of Various Agilent Sources
- Increase in Measured Noise as Ref Source Approaches DUT Noise
- Approximate Sensitivity of Delay Line Discriminator
- AM Calibration
- Voltage Controlled Source Tuning Requirements
- Tune Range of VCO for Center Voltage
- Peak Tuning Range Required by Noise Level
- Phase Lock Loop Bandwidth vs. Peak Tuning Range
- Noise Floor Limits Due to Peak Tuning Range
- Tuning Characteristics of Various VCO Source Options
- 8643A Frequency Limits
- 8644B Frequency Limits
- 8664A Frequency Limits
- 8665A Frequency Limits
- 8665B Frequency Limits
- System Specifications
- System Interconnections
- PC Components Installation
- Overview
- Step 1: Uninstall the current version of Agilent Technologies IO libraries
- Step 2: Uninstall all National Instruments products.
- Step 3: Install the National Instruments VXI software.
- Step 4: Install the National Instruments VISA runtime.
- Step 5: Install software for the NI Data Acquisition Software.
- Step 6: Hardware Installation
- Step 7. Finalize National Instruments Software Installation.
- Step 8: System Interconnections
- Step 9: Install Microsoft Visual C++ 2008 Redistributable Package use default settings
- Step 10: Install the Agilent I/O Libraries
- Step 11: Install the E5500 Phase Noise Measurement software.
- Step 12: Asset Configuration
- Step 13: License Key for the Phase Noise Test Set
- Overview
- PC Digitizer Performance Verification
- Preventive Maintenance
- Service, Support, and Safety Information
- Safety and Regulatory Information
- Safety summary
- Equipment Installation
- Environmental conditions
- Before applying power
- Ground the instrument or system
- Fuses and Circuit Breakers
- Maintenance
- Safety symbols and instrument markings
- Regulatory Compliance
- Declaration of Conformity
- Compliance with German noise requirements
- Compliance with Canadian EMC requirements
- Service and Support
- Return Procedure
- Safety and Regulatory Information
Residual Measurement Fundamentals
7
Agilent E5505A User’s Guide 215
Measured ± DC peak voltage
Advantages
• Easy method for calibrating the measurement system.
• This calibration technique can be performed using the baseband analyzer.
• Fastest method of calibration. If, for example, the same power levels are
always at the phase detector, as in the case of leveled, or limited outputs,
the phase detector sensitivity will always be essentially equivalent (within
one or two dB). Recalibration becomes unnecessary if this accuracy is
adequate.
• Only one RF source is required.
• Measures the phase detector gain in the actual measurement configuration.
This technique requires you to adjust off of quadrature to both the positive
and the negative peak output of the Phase Detector. This is done by either
adjusting the phase shifter or the frequency of the source. An oscilloscope
or voltmeter can optionally be used for setting the positive and negative
peaks.
Disadvantages
• Has only moderate accuracy compared to the other calibration methods.
• Does not take into account the amount of phase detector harmonic
distortion relative to the measured phase detector gain, hence the phase
detector must operate in its linear region.
• Requires manual adjustments to the source and/or phase shifter to find the
phase detector’s positive and negative output peaks. The system will read
the value of the positive and negative peak and automatically calculate the
mean of the peak voltages which is the phase detector constant used by the
system.
Procedure
1
Connect circuit as per Figure 152 on page 216, and tighten all connections.
2
Measure the power level that will be applied to the Signal Input port of the
test set’s phase detector. Table 30 on page 216 shows the acceptable
amplitude ranges for the E5505A system phase detectors.