User`s guide

Table Of Contents
Residual Measurement Fundamentals
7
Agilent E5505A User’s Guide 215
Measured ± DC peak voltage
Advantages
Easy method for calibrating the measurement system.
This calibration technique can be performed using the baseband analyzer.
Fastest method of calibration. If, for example, the same power levels are
always at the phase detector, as in the case of leveled, or limited outputs,
the phase detector sensitivity will always be essentially equivalent (within
one or two dB). Recalibration becomes unnecessary if this accuracy is
adequate.
Only one RF source is required.
Measures the phase detector gain in the actual measurement configuration.
This technique requires you to adjust off of quadrature to both the positive
and the negative peak output of the Phase Detector. This is done by either
adjusting the phase shifter or the frequency of the source. An oscilloscope
or voltmeter can optionally be used for setting the positive and negative
peaks.
Disadvantages
Has only moderate accuracy compared to the other calibration methods.
Does not take into account the amount of phase detector harmonic
distortion relative to the measured phase detector gain, hence the phase
detector must operate in its linear region.
Requires manual adjustments to the source and/or phase shifter to find the
phase detector’s positive and negative output peaks. The system will read
the value of the positive and negative peak and automatically calculate the
mean of the peak voltages which is the phase detector constant used by the
system.
Procedure
1
Connect circuit as per Figure 152 on page 216, and tighten all connections.
2
Measure the power level that will be applied to the Signal Input port of the
test set’s phase detector. Table 30 on page 216 shows the acceptable
amplitude ranges for the E5505A system phase detectors.