152 Agilent E5505A User’s Guide
6
Absolute Measurement Examples
3
When the Connect Diagram dialog box appears, click on the hardware
drop-down arrow and select your hardware configuration from the list. See
Figure 104.
4
Connect your DUT and reference sources to the test set at this time.
Confirm your connections as shown in the connect diagram.
The input attenuator (Option 001 only) is now correctly configured
based on your measurement definition.
Figure 104 Connect diagram for the free-running RF oscillator measurement
CAUTION
The test set’s signal input is subject to the limits and characteristics in
Tab le 1 8
on
page 152.
To prevent damage to the test set’s components, do not apply the input signal to
the test set’s signal input connector until the input attenuator (Option 001) has
been set by the phase noise software, which will occur when the connection
diagram appears.
Table 18 Test set signal input limits and characteristics
Limits