User`s guide

Table Of Contents
Agilent E5505A User’s Guide 15
Method 2, example 2 298
Figure 230. Double-sided spur AM noise setup (method 2, example 2) 298
Figure 231. Measuring power at the am detector 298
Figure 232. Measuring carrier-to-sideband ratio 299
Figure 233. Measuring the calibration constant 299
Method 3: Single-Sided Spur 301
Figure 234. AM noise measurement setup using single-sided spur 301
Figure 235. Measuring relative spur level 302
Figure 236. Measuring detector sensitivity 302
12 AM Noise Measurement Examples
AM Noise with N5500A Option 001 304
Required equipment 304
Figure 237. AM noise measurement configuration 304
Defining the measurement 304
Figure 238. Select the parameters definition file 305
Figure 239. Navigate to AM noise 306
Figure 240. Enter Frequencies in Source Tab 306
Figure 241. Enter parameters into the cal tab 307
Figure 242. Select parameters in the block diagram tab 307
Figure 243. Select graph description on graph tab 308
Beginning the measurement 308
Figure 244. Selecting a new measurement 308
Figure 245. Confirm measurement dialog box 309
Figure 246. Connect diagram for the AM noise measurement 309
Table 47. Test set signal input limits and characteristics 311
Figure 247. Connect diagram example 311
Making the measurement 312
When the measurement is complete 312
Figure 248. Typical AM noise curve 312
Table 48. Parameter data for the AM noise using an N5500A Option 001 313
13 Baseband Noise Measurement Examples
Baseband Noise with Test Set Measurement Example 316
Defining the measurement 316
Figure 249. Select the parameters definition file 316
Beginning the measurement 317
Figure 250. Selecting a new measurement 317
Figure 251. Confirm measurement dialog box 317
Figure 252. Connect diagram dialog box 318
Making the measurement 318