6
Absolute Measurement Examples
Selecting a reference source
1
Using Figure 86 on page 133 as a guide, navigate to the Block Diagram tab.
2
From the Reference Source pull-down list, select your source.
Table 14 Tuning characteristics for various sources
VCO Source Carrier
Freq.
Tuning Constant
(Hz/V)
Center
Voltage
(V)
Voltage Tuning
Range (±V)
Input
Resistance
(Ω)
Tuning
Calibration
Method
Agilent 8662/3A
EFC
DCFM
υ
0
5 E – 9 x υ
0
FM Deviation
0
0
10
10
1E + 6
1 K (8662)
600 (8663)
Measure
Compute
Compute
Agilent 8642A/B FM Deviation 0 10 600 Compute
Agilent 8644B FM Deviation 0 10 600 Compute
Other Signal
Generator
DCFM Calibrated for
±1V
FM Deviation 0 10 R
in
Compute
Other User VCO
Source
Estimated within a
factor of 2
–10 to
+10
1 E + 6 Measure