Absolute Measurement Fundamentals
5
Agilent E5505A User’s Guide 115
Estimating the Tuning Constant
The VCO tuning constant is the tuning sensitivity of the VCO source in Hz/V.
The required accuracy of the entered tuning constant value depends on the
VCO tuning constant calibration method specified for the measurement. The
calibration method is selected in the Calibration Process menu. Table 13 lists
the calibration method choices and the tuning constant accuracy required for
each.
Table 13 VCO tuning constant calibration method
VCO Tuning Constant Calibration Method (selected in
calibration screen)
Required Tuning Constant Accuracy
(entered in parameter screen)
Use the current tuning constant
(must be accurate from a previous measurement of the
same source).
Within a factor of 2 of actual value.
(Enter 1 E + 6 for Input Resistance.)
Measure the VCO tuning constant Within a factor of 2 of actual value.
(Enter 1 E + 6 for Input Resistance.)