Service manual

C
c
16.
Press MEASUREMENT
IRESTART).
17.
Press MENUS
(a)
and use the front panel knob or the
numeric keypad to set the marker to the closest point to the first
frequency listed in the test record at the end of this chapter.
18.
Divide the values shown on the analyzer by two and record this
value in
the test record.
The value shown on the analyzer represents an out-and-back
path of the signal, which is twice the cable insertion loss
19. Repeat the previous two steps for the remaining values on the
test record.
HP
85133E/F
Performance Tests 3-5