User manual

1
Instructions for Use
Adjustment of Handle Position
20/36 Variable Leak Valve User Manual / 87-400-085-01 (D)
Adjustment of Handle Position
As described previously, continual reduction in gasket height
necessitates resetting of the stop position and eventual readjustment
of handle position. When the angle between the handle and body at
point of closure has changed from its original parallel (approximate)
position to an angle of +/-5°, the valve should be readjusted. This
adjustment will maintain a proper relationship between handle travel
and the size of the leak.
1. Set handle.
Turn the locked knobs counterclockwise until the valve handle is
approximately parallel to the side of the valve body.
2. Close valve using roughing screw as follows.
a Carefully remove the hole cover from the top of the valve.
CAUTION!
Do not exceed 8 ft.-lbs. or torque or the sapphire may be fractured.
b With a 5/16” Allen wrench, tighten the roughing screw to 6 ft.-lbs.
of torque.
c Replace the hole cover.
d The valve should now be leak-tight, but this can be determined
reliably only with a leak detector. If the valve leaks across the
seal, refer to the following paragraph.
e Adjust knobs as follows.
f Turn the knobs together counterclockwise until a leak is
generated.
g Turn the knobs clockwise until the valve is closed leak-tight.
h Turn the knobs two additional turns clockwise to provide the
proper amount of overdrive.
i Loosen the knobs without allowing the fine-drive screw to turn
then turn the inner knob clockwise until the collar is against the
handle. Lock the outer knob against the inner one.
j Two counterclockwise turns of both knobs together should open
a leak.