User manual
4
Instructions for Use
Use Procedure
46/70 MiniVac Controller User Manual / 87-900-049-01 (H)
Use Procedure
Interlock
The Interlock is a safety device designed by Agilent to protect the
operator against serious injury caused by the high voltage present on
the pump power connector.
The controller is intended to be used with a HV cable furnished with
Interlock.
If the customer does not wish to use the HV cable with interlock for
particular application requirements, the interlock cable of the
controller (4) is available.
Start-Up of the Pump
To start the pump:
1. Connect the pump to the controller using the specific HV cable.
2. Connect the Interlock of the HV cable to the plug (5); if the HV
cable is without the interlock cable, connect the interlock cable of
the controller (4) to the plug (5).
3. Switch on the controller.
Controller Start-up
To start-up the controller, turn the main switch to ON, after plugging
the power cable into a suitable power source.