Technical data

Troubleshooting 4
6000 Series Oscilloscopes Service Guide 87
To verify basic oscilloscope operation
For 6000L Series oscilloscopes, connect the oscilloscope to
your network or establish a point- to- point connection, and
use the Remote Front Panel feature described in the 6000
Series Oscilloscopes User’s Guide to control the oscilloscope.
If you do not have the user’s guide, you can download it
from www.agilent.com/find/mso6000.
1 Press the Save/Recall key on the front panel, then press
the Default Setup softkey under the display. The oscilloscope
is now configured to its default settings.
2 Connect an oscilloscope probe from channel 1 to the Probe
Comp signal terminal on the front panel.
3 Connect the probe’s ground lead to the ground terminal
that is next to the Probe Comp terminal.
4 Press AutoScale.
5 You should see a waveform on the oscilloscope’s display
similar to this: