Technical data
Testing Performance 2
6000 Series Oscilloscopes Service Guide 71
Horizontal ∆t Accuracy
Generator Setting Test Limits Results
Period 100 µs 99.8 µs to 100.2 µs ________
Period 100 ns 99.8 ns to 100.2 ns ________
Period 10 ns
1
9.93 ns to 10.07 ns ________
Period 5 ns
2
4.96 ns to 5.04 ns ________
1
100 MHz models
2
300 MHz, 500 MHz, and 1 GHz models
Trigger Sensitivity Test Limits Channel 1 Channel 2 Channel 3* Channel 4*
Internal trigger
1 division at 25 MHz ________ ________ ________ ________
1 GHz models 0.6 division at 1 GHz ________ ________ ________ _______
500 MHz models 0.6 division at 500 MHz ________ ________ ________ _______
300 MHz models 0.6 division at 300 MHz ________ ________ ________ _______
100 MHz models 0.6 division at 100 MHz ________ ________ ________ _______
Trigger Sensitivity Test Limits Channel 1 Channel 2 Channel 3* Channel 4*
External trigger (6102A, 6052A, 6032A, 6012A)
± 1 V range:
100 mV at 25 MHz ________ ________ ________ ________
6102A 200 mV at 1 GHz ________ ________ ________ ________
6052A 200 mV at 500 MHz ________ ________ ________ ________
6032A 200 mV at 300 MHz ________ ________ ________ ________
6012A 200 mV at 100 MHz ________ ________ ________ ________
External trigger (6102A, 6052A, 6032A, 6012A)
± 8 V range:
250 mV at 25 MHz ________ ________ ________ ________
6102A 500 mV at 1 GHz ________ ________ ________ ________
6052A 500 mV at 500 MHz ________ ________ ________ ________
6032A 500 mV at 300 MHz ________ ________ ________ ________
6012A 500 mV at 100 MHz ________ ________ ________ ________
External trigger (6014, 6034A, 6054, 6104)
6104, 6054, 6034 500 mV at 500 MHz ________ ________ ________ ________
6014 500 mV at 100 MHz ________ ________ ________ ________
* Where applicable
6000_series_service_guide.book Page 71 Friday, March 23, 2007 2:43 PM