Technical data
70 6000 Series Oscilloscopes Service Guide
2 Testing Performance
Agilent 6000 Series Oscilloscopes Performance Test Record
Serial No. ______________________________________ Test by _____________________________
Test Interval ____________________________________ Work Order No. ______________________
Recommended Next Testing ________________________ Temperature ____________
Threshold Specification Limits Ch D7-D0 Ch D15-D8
Accuracy Test 5 V - 250 mV 4.750 V ________ ________
(100 mV + 3% of
threshold setting)
5 V + 250 mV 5.250 V ________ ________
-5 V - 250 mV -5.250 V ________ ________
-5 V + 250 mV -4.750 V ________ ________
0 V - 100 mV -100 mV ________ ________
0 V + 100 mV 100 mV ________ ________
Voltage Measurement Accuracy
Range Power Supply Setting Test Limits Channel 1 Channel 2 Channel 3* Channel 4*
5 V/Div 35 V 34.04 V to 35.96 V ________ ________ ________ ________
2 V/Div 14 V 13.616 V to 14.384 V ________ ________ ________ ________
1 V/Div 7 V 6.808 V to 7.192 V ________ ________ ________ ________
500 mV/Div 3.5 V 3.404 V to 3.596 V ________ ________ ________ ________
200 mV/Div 1.4 V 1.3616 V to 1.4384 V ________ ________ ________ ________
100 mV/Div 700 mV 680.8 mV to 719.2 mV ________ ________ ________ ________
50 mV/Div 350 mV 340.4 mV to 359.6 mV ________ ________ ________ ________
20 mV/Div 140 mV 136.16 mV to 143.84 mV ________ ________ ________ ________
10 mV/Div 70 mV 68.08 mV to 71.92 mV ________ ________ ________ ________
5 mV/Div 35 mV 34.04 mV to 35.96 mV ________ ________ ________ ________
2 mV/Div
(300 MHz, 500 MHz, and 1 GHz models)
14 mV 13.232 mV to 14.768 mV ________ ________ ________ ________
2 mV/Div
(100 MHz models)
14 mV 13.616 mV to 14.384 mV ________ ________ ________ ________
1 mV/Div
(MSO/DSO6012A)
7 mV 6.616 mV to 7.384 mV ________ ________ ________ ________
Bandwidth Model Test Limits Channel 1 Channel 2 Channel 3* Channel 4*
610x 3 dB at 1 GHz ________ ________ ________ ________
605x 3 dB at 500 MHz ________ ________ ________ ________
603x 3 dB at 300 MHz ________ ________ ________ ________
601x 3 dB at 100 MHz ________ ________ ________ ________
6000_series_service_guide.book Page 70 Friday, March 23, 2007 2:43 PM