Technical data
62 6000 Series Oscilloscopes Service Guide
2 Testing Performance
Test Internal Trigger Sensitivity (all models)
1 Connect the equipment (see Figure 6).
a For 300 MHz, 500 MHz, and 1 GHz models, connect the
signal generator output to the oscilloscope channel 1
input.
b For 100 MHz models, use a 50Ω feedthrough
termination to connect the signal generator output to
the oscilloscope channel 1 input.
Figure 6 Connect equipment for internal trigger sensitivity test
Signal
Generator
N to BNC adapter
N Cable
50Ω Feedthrough*
* 100 MHz models only.
Oscilloscope