Technical data
Testing Performance 2
6000 Series Oscilloscopes Service Guide 61
Test limits for:
• Internal trigger on all models:
• < 10 mV/div: greater of 1 div or 5 mV
pp
>=10 mV/div: 0.6 div
• External trigger on all 2- channel models
(DSO/MSO6xx2A):
• Trigger range: 1.0V
DC to 100 MHz: < 100 mV
pp
>100 MHz to max bandwidth: < 200 mV
pp
• Trigger range: 8.0V
DC to 100 MHz: < 250 mV
pp
>100 MHz to max bandwidth: < 500 mV
pp
• External trigger on 4- channel models:
4- channel 1 GHz, 500 MHz, and 300 MHz models:
(MSO/DSO6104A/6054A/6034A, DSO6104L/6054L)
DC to 500 MHz: < 500 mV
pp
4- channel 100 MHz models:
(MSO/DSO6014A, DSO6014L)
DC to 100 MHz: < 500 mV
pp
Table 12 Equipment Required to Verify Trigger Sensitivity
Equipment Critical Specifications Recommended Model/Part
Signal
Generator
25-MHz, 100-MHz, 300-MHz,
500-MHz, and 1 GHz sine
waves
Agilent E4400B/8648A
Power splitter Outputs differ < 0.15 dB Agilent 11667A
Cable BNC, Qty 3 Agilent 10503A
Adapter N (m) to BNC (f), Qty 3 Agilent 1250-0780
Feedthrough 50Ω BNC (f) to BNC (m) Qty 2 Agilent 0960-0301