Technical data
52 6000 Series Oscilloscopes Service Guide
2 Testing Performance
The ∆Y value on the lower line of the display should be
within the test limits of Table 9. If a result is not
within the test limits, go to the “Troubleshooting”
chapter. Then return here.
6 Continue to check the voltage measurement accuracy with
the remaining Volts/div setting values in Table 9.
7 When you are finished checking all of the voltage values,
disconnect the oscilloscope calibrator from the
oscilloscope.
8 Repeat this procedure for the remaining channels to be
tested.
Figure 4 Using a Blocking Capacitor to Reduce Noise
Use a Blocking Capacitor to Reduce Noise
On the more sensitive ranges, such as 2 mV/div and 5 mV/div, noise may be a factor. To
eliminate the noise, add a BNC Tee, blocking capacitor, and BNC shorting cap at the
oscilloscope channel input to shunt the noise to ground. See Figure 4.
To oscilloscope input
BNC shorting
cap
Blocking
Capacitor
6000_series_service_guide.book Page 52 Friday, March 23, 2007 2:43 PM