Technical data

Testing Performance 2
6000 Series Oscilloscopes Service Guide 51
3 Use the BNC tee and cables to connect the oscilloscope
calibrator /power supply to both the oscilloscope and the
multimeter (see Figure 3).
Figure 3 Connect equipment for voltage measurement accuracy test
4 Adjust the output so that the multimeter reading displays
the first Volts/div calibrator setting value in Table 9.
Wait a few seconds for the measurement to settle.
5 Press the Y2 softkey, then position the Y2 cursor to the
center of the voltage trace using the Entry knob.
Oscilloscope
Oscilloscope
Calibrator
Digital
Multimeter
BNC Tee
BNC (f) to dual
banana
6000_series_service_guide.book Page 51 Friday, March 23, 2007 2:43 PM