Technical data

50 6000 Series Oscilloscopes Service Guide
2 Testing Performance
c Press the Acquire key. Then press the Averaging softkey
and set #Avgs to 64.
Wait a few seconds for the measurement to settle.
2 Press the Cursors key, set the Mode softkey to Normal, then
press the XY softkey and select Y. Press the Y1 softkey,
then use the Entry knob (labeled on the front panel)
to set the Y1 cursor on the baseline of the signal.
Table 9 Settings Used to Verify Voltage Measurement Accuracy
Volts/Div Setting Oscilloscope
Calibrator Setting
Test Limits
5 V/Div 35 V 34.04 V to 35.96 V
2 V/Div 14 V 13.616 V to 14.384 V
1 V/Div 7 V 6.808 V to 7.192 V
500 mV/Div 3.5 V 3.404 V to 3.596 V
200 mV/Div 1.4 V 1.3616 V to 1.4384 V
100 mV/Div 700 mV 680.8 mV to 719.2 mV
50 mV/Div 350 mV 340.4 mV to 359.6 mV
20 mV/Div 140 mV 136.16 mV to 143.84 mV
10 mV/Div 70 mV 68.08 mV to 71.92 mV
5 mV/Div 35 mV 34.04 mV to 35.96 mV
2 mV/Div
1
(for 1 GHz, 500 MHz, and 300 MHz models)
14 mV 13.232 mV to 14.768 mV
2 mV/Div (for 100 MHz models)
14 mV 13.616 mV to 14.384 mV
1 mV/Div
2
7 mV 6.616 mV to 7.384 mV
1
Full scale is defined as 32 mV on the 2 mV/div range for 300 MHz, 500 MHz, and 1 GHz models.
2
Full scale is defined as 16 mV on the 1 mV/div range for 100 MHz models.
Full scale on all other ranges is defined as 8 divisions times the V/div setting.