Technical data
Testing Performance 2
6000 Series Oscilloscopes Service Guide 49
1 Set up the oscilloscope.
a Adjust the channel 1 position knob to place the
baseline at 0.5 major division from the bottom of the
display.
b Set the Volts/Div setting to the value in the first line in
Table 9.
Table 8 Equipment Required to Verify Voltage Measurement Accuracy
Equipment Critical Specifications Recommended
Model/Part
Oscilloscope
Calibrator
14 mV to 35 Vdc,
0.1 V resolution
Fluke 5820A
Digital multimeter Better than 0.01% accuracy Agilent 34401A
Cable BNC, Qty 2 Agilent 10503A
Shorting cap BNC Agilent 1250-0774
Adapter BNC (f) to banana (m) Agilent 1251-2277
Adapter BNC tee (m) (f) (f) Agilent 1250-0781
Blocking capacitor Agilent 10240B