Technical data

46 6000 Series Oscilloscopes Service Guide
2 Testing Performance
Figure 2 Setting Up Equipment and Test Connector for the Threshold
Te s t
4 Use a BNC-banana cable to connect the multimeter to the
other side of the BNC Tee.
5 Connect the BNC Tee to the Channel 1 output of the
calibrator as shown in Figure 2.
6 On the oscilloscope, press the D15 Thru D0 key, then press
the Thresholds softkey, then press the D15 Thru D0 softkey
repeatedly until the check mark is next to User.
Test
Connector
BNC-Banana
cable
Channels
8 - 15
Channels
0 - 7
Oscilloscope
Calibrator
Digital
Multimeter
BNC Tee
Probe
Cables
Channels
8 - 15
Channels
0 - 7
6000A Series Oscilloscope
6000L Series Oscilloscope
6000_series_service_guide.book Page 46 Friday, March 23, 2007 2:43 PM