Technical data

40 6000 Series Oscilloscopes Service Guide
2 Testing Performance
List of Test Equipment
Below is a list of test equipment and accessories required to
perform the performance test verification procedures.
Table 4 List of test equipment
Equipment Critical Specifications
Recommended Model/
Part Number
Test connector, 8-by-2* See page 41 for instructions on building test connector. n/a
Digital Multimeter 0.1 mV resolution, 0.005% accuracy Agilent 34401A
Power Splitter Outputs differ by 0.15 dB Agilent 11667B
Oscilloscope Calibrator DC offset voltage of -5.5 V to 35.5 V, 0.1 V resolution Fluke 5820A
Signal Generator 25 MHz, 100 MHz, 300 MHz, 500 MHz, and 1 GHz
sine waves
Agilent E4400B or
Agilent 8648A
Power Meter/Sensor 1 GHz
±3% accuracy Agilent E4418B/8482A
Oscilloscope Calibrator 25 MHz—500 MHz sine wave, 5 ppm Fluke 5820A
BNC banana cable Agilent 11001-66001
BNC cable (qty 3) Agilent 10503A
Cable Type N (m) 609.6 mm (24 in.) Agilent 11500B
Probe cable* Agilent 01650-61607
Shorting Cap BNC Agilent 1250-0774
Adapter BNC(f) to banana(m) Agilent 1251-2277
Adapter BNC Tee (m) (f) (f) Agilent 1250-0781
Adapter Type N (m) to BNC (m) Agilent 1250-0082
or Pomona 3288 with
Pomona 3533
Blocking capacitor Agilent 10240-60001
Adapter (qty 3) N(m) to BNC(f) Agilent 1250-0780
Feedthrough (qty 2) 50
BNC (f) to BNC (m) Agilent 0960-0301
* Required only for testing digital channels of oscilloscopes that have the MSO option.
Some parts and equipment are available at www.agilent.com or www.parts.agilent.com.