Technical data

Replacing 6000L Assemblies 6
6000 Series Oscilloscopes Service Guide 153
4 Replace the fan.
a Place the rectangular part of the flexible fan mount
over the fan.
b Place the pegs in the flexible fan mount into the holes
in the top and bottom of the fan.
c Align the holes in the bottom of the flexible fan mount
over the studs on the chassis as you slide the fan over
the divider in the chassis.
d Ensure that the notch in the rectangular portion of the
fan mount aligns with the front of the chassis.
Figure 55 Installing the fan.
Pegs Fit in
Fan Holes
Holes in
Bottom Fit
Over Studs
Align
Notch
and
Chassis
Chassis
Divider
6000_series_service_guide.book Page 153 Friday, March 23, 2007 2:43 PM