Technical data

146 6000 Series Oscilloscopes Service Guide
6 Replacing 6000L Assemblies
To remove the front panel assembly
1 Perform the following procedures:
“To remove the bottom cover” on page 145
2 Insert a flat- blade screwdriver under the center of the
intensity knob and gently twist the screwdriver as you
pull the knob off.
Using a twisting motion on the screwdriver rather than
prying prevents marking or damaging the front panel.
Figure 49 Removing the intensity knob
3 Disconnect the front panel ribbon cable from the system
board by pressing the release tabs and then disconnect it
from the front panel PC board.
4 Use a flat- blade screwdriver to release the center retainer
tab and push it slightly out of the slot.
5 Use a flat- blade screwdriver to release the 2 side retainer
tabs and remove them from their slots.
Intensity
Knob
6000_series_service_guide.book Page 146 Friday, March 23, 2007 2:43 PM