Technical data
Replacing 6000A Assemblies 5
6000 Series Oscilloscopes Service Guide 139
To remove the system board
1 Perform the previous procedures:
• “To remove the cabinet” on page 104
2 Remove the main shield covering the system board by
sliding it toward the back of the instrument.
3 Using a small flat blade screw driver, gently pry and
remove the display cable and backlight inverter cable.
4 Disconnect the fan cable and keyboard ribbon cable then
push all cables back through the hole in the system board.
WARNING
Thin sheet metal parts may have sharp edges. Handle with care to avoid injury.
6000_series_service_guide.book Page 139 Friday, March 23, 2007 2:43 PM