Technical data

Replacing 6000A Assemblies 5
6000 Series Oscilloscopes Service Guide 137
To remove the fan
1 Perform the previous procedures:
“To remove the cabinet” on page 104
“To remove the power supply shield” on page 125
“To remove the power supply” on page 129
2 Remove the main shield covering the system board by
sliding it toward the back of the instrument.
3 Disconnect the fan cable from the system board.
Figure 44 Removing main shield and disconnecting fan cable.
WARNING
Thin sheet metal parts may have sharp edges. Handle with care to avoid injury.
Main shield
Fan cable