Technical data

124 6000 Series Oscilloscopes Service Guide
5 Replacing 6000A Assemblies
5 To reinstall:
a Place the protective lens into the pocket in the display
mount and ensure that the inside of the lens is clean.
b Place the gasket into the slot around the lens making
sure it is fully seated in the slot all the way around.
c Clean the LCD window and insert the LCD under the
guides on the right side of the display mount.
d Push the left side of the LCD down until it clicks under
the latches so that the latches are fully over the face of
the sheet metal housing.
Figure 33 Latch over face of sheet metal housing
e Reinstall the backlight inverter (see step 9 on
page 120).
Latch
Sheet metal face
6000_series_service_guide.book Page 124 Friday, March 23, 2007 2:43 PM