Technical data
122 6000 Series Oscilloscopes Service Guide
5 Replacing 6000A Assemblies
To remove the LCD, gasket, and protective lens from the display
mount
1 Perform the following procedures:
• “To remove the cabinet” on page 104
• “To remove the front panel assembly” on page 107
• “To remove the display assembly” on page 113
• “To remove the backlight inverter board and e-field
shield” on page 117
2 Use a long- nose pliers or flat- blade screwdriver to push
and release the two LCD latches on the left side of the
display mount.
Figure 31 Release display mount latches
LCD Latches
Display Mount
LCD
6000_series_service_guide.book Page 122 Friday, March 23, 2007 2:43 PM