Technical data
Replacing 6000A Assemblies 5
6000 Series Oscilloscopes Service Guide 121
f Close the shield.
- Insert the shield between the plastic and sheet metal
of the display.
- Ensure the cables do not get pinched.
- After closing, use your fingers to gently pinch along
the perforated edge of the shield to square the corner
and ensure that the release tabs are in their slots.
Figure 30 Closing the shield
Perforated Edge
Plastic
Display
Sheet Metal