Technical data

Replacing 6000A Assemblies 5
6000 Series Oscilloscopes Service Guide 113
To remove the display assembly
1 Perform the previous procedures:
“To remove the cabinet” on page 104
“To remove the front panel assembly” on page 107
2 Remove the main shield covering the system board by
sliding it toward the back of the instrument.
3 Using a small flat blade screw driver, gently pry the
display and backlight inverter cables to disconnect them
from the system board.
Note the cable routing through the system board for
reinstallation.
WARNING
Thin sheet metal parts may have sharp edges. Handle with care to avoid injury.