Service manual
98 Replaces page 37
Table 3-12. Transient Generator Signature Analysis (Secondary)
Description:
These signatures check transient generator IC's U310 through U316. The signatures are valid for U301
firmware revisions "Rev A.01.01" and "Rev A.01.02". Use the test setup described in the "Test Setup for S.A.".
Signatures:
+5V = 1CAU U312-16 U313-10,16 U310-16 U311-16 U314-10,15,16 U316-14
Common U312-8 U313-8 U310-8 U311-8 U314-3,8 U316-7
SD(0) U312-15 = lP45 U313-15 = lP45
SD(1) U312-1 = 200F U313-1= 2OOF
SD(2) U312-2 = 89AA U313-2 = 89AA
SD(3) U312-3 = 93AU U313-3 = 93AU
SD(4) U312-4 = 7792 U313-4 = 7792
SD(5) U312-5 = 0343 U313-5 = 0343
SD(6) U312-6 = FPOU U313-6 = FPOU
SD(7) U312-7 = 6740 U313-7 = 674O
U312-9 = 1CAU pulsing U313-9 = 84U8
U312-10 = 1CAU pulsing
U312-11 = 433A U313-11 = 433A
U312-12 = 707A U313-12 = 6CH5
U312-13 = 48U7 U313-13 = 6746
U312-14 = 84U8 U313-14 = 1CAU pulsing
U310-10 = 1 MHz U314-1 = 1CAU U314-9 = 6CH5
U310-11 = OOOO pulsing U314-2 = 9AFP U314-10 = 1CAU +5V
U314-3 = OOOO U314-11 = 9U57
U311-9 = OOOU U314-4 = 1CAU pulsing U314-U - OOOO pulsing
U311-10 = lUF1 U314-5 = 1CAU U314-13 = 5895
U311-11 = 83OO U314-6 = OOOO U314-14 = 1CAU pulsing
U314-7 = 707A U314-15 = 1CAU +5V
U315-1 = 1CAU pulsing U316-1 = 84U8 U316-8 = 9AFP
U315-2 = OOOO pulsing U316-2 = OOOO U316-9 = 84U8
U315-10 = OOOO pulsing U316-3 = 9U57 U316-10 = 9U57
U315-11 = 1CAU pulsing U316-4 = 9U57 U316-11 = OOOO pulsing
U316-5 = 84U8 U316-12 = OOOO pulsing
U316-6 = 84U8 U316-13 = 1CAU pulsing