Operating instructions

Status Reporting
88
Table 5-3. Standard Event Status Bit Description (continued)
Mnemonic Bit
1
Value Meaning
CME 5 32 Command Error. A syntax or semantic error has occurred or the electronic
load received a < GET > within a program message. Errors in the range of
-199 thru -l00 can set this bit.
PON 7 128 Power On. The electronic load has been turned on or off since the last time this
register was read. This bit is always set when the electronic load is turned on.
1
Bits 1 and 7 are not used by the electronic load.
Operation Status
The Operation Status registers let you determine whether either of the operation conditions described in Table 5-4
presently exist on the electronic load.
The Operation Condition register represents the present status of the electronic load; the bits are set when the indicated
condition is true.
The PTR/NTR filter determines what type of transition in the Operation Condition register will set the bit in the
corresponding Operation Event register.
Programming a bit in the PTR filter causes a 0-to-1 transition in the Operation Condition register to set the
corresponding bit in the Operation Event register.
Programming the NTR filter causes a 1-to-0 transition to set the bit.
Programming both filters causes either transition to set the bit.
If the transition filters are not programmed, the bit is disabled.
Table 5-4. Operation Status Bit Description
Mnemonic Bit
1
Value Meaning
CAL 0 1 Calibrating. A calibration calculation is in progress. (Refer to the Operating
Manual for details of calibration commands.)
WTG 5 32 Waiting for trigger. At least one channel is waiting for a trigger to occur. Any
TRIG command for any mode on any channel sets this bit. When a trigger is
received, the bit is reset. ABORt also resets this bit.
1
Bits 1-4, and 6-15 are not used by the electronic load.
Note When the unit is turned on, the PTR filter is programmed on for the CAL bit, and the NTR filter is
programmed on for the WTG bit.
The Operation Event register represents all of the filtered operation conditions that have occurred since the last time this
register was read. Reading the register resets it to zero.
The Operation Enable register can be programmed to specify which operation event bits are logically-ORed to become Bit
7 (OPER bit) in the Status Byte register.
Note Refer to the Electronic Load Operating Manual for information about calibration.