Operating instructions
Language Dictionary
75
TRIG:TIMer Channel-Independent Trigger Command/Query
Description
This command determines the period of the trigger pulses generated by the multiple
electronic load’s internal trigger oscillator. The trigger oscillator begins running as soon as
this command is executed. If the command is executed when the TRIG:SOUR is not
TIMer, the programmed period will take effect when the next TRIG:SOUR:TIM
statement is executed. If the TIM parameter is out of the specified limits, an error is
generated (See Table 4-2 at the end of this chapter).
Note
Do not confuse this command with TRIG:SOUR TIM. In that command, TIM is a
parameter for SOURce. This TIM is on the same tree level as SOURce and has its own
parameter.
Command Syntax
TRIGger:TIMer <NRf+>
Parameters
See Table 4-1 and the Operating Manual of the electronic load model.
Examples
TRIG:SOUR TIM;TIM .001 Select time mode with frequency = 1 kHz.
TRIG: TIM 5E-5 Change trigger frequency to 20 kHz.
Query Syntax
TRIG:TIM?
Returned Parameters
<NR3> Value representing period in seconds.
Related Commands
None
VOLTage Subsystem Channel-Specific Voltage Programming Function
Description
This subsystem programs the voltage-mode function of a single electronic load or a single
channel of a multiple electronic load.
Keywords
Command
VOLTage[:LEVel][:IMMediate]
Function
Specify main level for CV Mode.
VOLTage[:LEVel]:TRlGgered
Preset voltage level pending trigger occurrence.
VOLTage:SLEW
Specify voltage level rate of change.
VOLTage:TLEVel
Specify input voltage level used by TRANsient subsystem.
Related Subsystems
CURRent RESistance