Operating instructions

Language Dictionary
72
TRlGger Subsystem Channel-Independent Trigger Programming Function
Description
In a multiple electronic load, triggering can be initiated from the controller, the a-c line,
the external Trigger jack, or the mainframe timer. single electronic loads may be triggered
from their external Trigger jack, or from the controller. Any trigger signal generated by a
single electronic load or multiple electronic load also exists at the Trigger output jack and
may be used as a trigger source for any other electronic load.
Note
For a detailed explanation of the effects of triggers, see Operation Overview in the
electronic load Operating Manual.
Keywords
Command
TRIGger[:IMMediate]
Function
Trigger-true input from the controller.
TRIGger:SOURce EXTernal
Selects trigger source at external Trigger jack.
TRIGger:SOURce BUS
Selects GPIB *TRG or <GET> as trigger source.
TRIGger:SOURce HOLD
Disables all trigger inputs except the TRlG:IMM command.
1
TRIGger:SOURce LINE
Selects the internal a-c line frequency detector as the trigger source.
1
TRIGger:SOURce TIMer
Selects internal trigger oscillator as the trigger source.
1
These commands apply only to multiple electronic loads
Related Commands
*TRG <GET>
Related Subsystems
TRANsient Subsystem