Operating instructions
Language Dictionary
54
Examples
PORT0 ON Set the port output high (logical 1)
CHAN 1;: PORT0 ON;:CHAN 2;:PORT0 OFF Set Channel 1 port high
and Channel 2 port low.
Query Syntax
PORT0?
Returned Parameters
<NR1> 1 for ON or 0 for OFF
Related Commands
None
RESistance Subsystem Channel-Specific Resistance Programming Function
Description
This subsystem programs the constant-resistance (CR) mode function of a single electronic
load or a single channel of a multiple electronic load. There is no SLEW command in this
subsystem. In the lowest resistance range the level changes in Volts/sec at the rate specified
by the present VOLT:SLEW command. The middle and high ranges change level in
Amps/sec at the rate specified by the present CURR:SLEW command.
Keywords
Command
RESistance[:LEVel][:IMMediate]
RESistance[:LEVel]:TRIGgered
RESistance:RANGe
RESistance:TLEVel
Function
Specify input resistance for CR mode.
Preset a new input resistance to be valid when a trigger occurs.
Specify full-scale input resistance range.
Specify input resistance used with TRANsient Subsystem.