Operating instructions

Language Dictionary
53
Parameters
Enter the desired mode as a string variable in either the full or abbreviated format. There
are no units. The *RST Default is CURRent.
Examples
MODE :CURR Set electronic load input to CC mode
CHAN 1;: MODE:CURR; :CHAN 2; :MODE:RES Set Channel 1 input to CC mode
and Channel 2 input to CR mode
Query Syntax
MODE?
Returned Parameters
<aard> String value: CURR, RES, or VOLT
Alternate Syntax
FUNCtion can be used as an alias for MODE.
Syntax Diagram
Mode Command Syntax Diagram
PORT0 Channel-Specific Output Command/Query
Description
This command controls the general-purpose digital port on the rear of each electronic load
or each channel of the multiple electronic load. The state of the port is with respect to the
common side of the electronic load input. Setting a port to the OFF or 0 state connects it
to the common; setting a port to the ON or 1 state disconnects it from common and sets it
to a TTL logical high.
The command is the same for both single and multiple electronic loads. To control a port
in a multiple electronic load you must first select the channel and then use PORT0 on that
channel.
Command Syntax
PORT0[:STATe] <NRf+>
Note
This command does not have a four-character short form. You must always include the
fifth character (zero).
Parameters
Value Range Units *RST Default
OFF or 0
ON or 1
None
None
0