Command Syntax
INPut:SHORt[:STATe] <NRf+>
Parameters
Value Range Units *RST Default
OFF or 0
ON or 1
None
None
0
Examples
INP:SHOR ON Set shorted input condition
INP:SHOR 1
CHAN 2;INP:SHOR 1;:CHAN 1;INP:SHOR 0
On multiple electronic load, short Channel 1 and unshort Channel 2
Query Syntax
INP:SHOR?
Returned parameters
<NR1> Value; 0 for unshorted, 1 for shorted
Related Commands
INP[:STAT]
INP[:STATe] Channel-Specific Input Command/Query
Description
This implied keyword turns the electronic load input ON and OFF. When INPut is OFF,
the electronic load will draw minimal current at its input. INP[STAT]e:OFF overrides
INPput:SHORt:ON. The presently programmed slew rate setting is not used when turning
the input on or off; the current and voltage change at their maximum rates .
Command Syntax
INPut [:STATe] <NRf+>
Parameters
Value Range Units *RST Default
OFF or 0
ON or 1
None
None
0
Examples
INP:STAT OFF Turn electronic load input off
INP 0
INP:STAT OFF;SHOR ON;:INP ON Turn input off, short the input, and turn input on.
Query Syntax
INP?
Returned Parameters
<NR1> Value; 0 for OFF, 1 for ON
Related Commands
CURR:PROT INP:SHOR
MEASure Channel-Specific Measurement Query
Description
This function consists of queries that return the current, voltage, and power at the input of
the electronic load.
Query Syntax
Query
MEASure:CURRent[:DC]?
MEASure:POWer[:DC]?
MEASure:VOLTage[:DC]?
Value Returned
electronic load input current
Computed electronic load input power
electronic load input voltage