Operating instructions

Language Dictionary
33
*IDN? Identification Query
Type
System Interface
Description
This query requests the electronic load to identify itself.
Query Syntax
*IDN?
Returned Parameters
<aard> form consisting of four strings separated by commas. The content of each string
is:
String Information
Agilent Technologies Manufacturer
xxxxA Four-digit model number followed
by a letter suffix
0 Always returns zero
a.xx.xx Revision level of primary
interface firmware
Example
Agilent Technologies,6060A,0,A.01.02 This identifies an Agilent Model 6060A
Electronic
Load; with primary interface firmware
revision A.01.02
Related Commands
*OPT *RDT?
*OPC Operation Complete Event Bit Command
Type
Device Status
Description
This command causes Bit 0 of the Standard Event Status Event register to be set when the
electronic load has completed all pending operations. (See *ESR? for the bit configuration
of this register.) Pending operations are complete when:
All previous commands have been executed.
Any change in the input level caused by previous commands has been completed.
(Effects of slew rate have been accounted for.)
No pending trigger level operations are set for the single electronic load or for any
channel of the multiple electronic load.
*OPC does not prevent processing of subsequent commands but Bit 0 will not be set until
all pending operations are complete.
Command Syntax
*OPC
Parameters
None
Related Commands
*WAI *OPC?