Operating instructions
Language Dictionary
33
*IDN? Identification Query
Type
System Interface
Description
This query requests the electronic load to identify itself.
Query Syntax
*IDN?
Returned Parameters
<aard> form consisting of four strings separated by commas. The content of each string
is:
String Information
Agilent Technologies Manufacturer
xxxxA Four-digit model number followed
by a letter suffix
0 Always returns zero
a.xx.xx Revision level of primary
interface firmware
Example
Agilent Technologies,6060A,0,A.01.02 This identifies an Agilent Model 6060A
Electronic
Load; with primary interface firmware
revision A.01.02
Related Commands
*OPT *RDT?
*OPC Operation Complete Event Bit Command
Type
Device Status
Description
This command causes Bit 0 of the Standard Event Status Event register to be set when the
electronic load has completed all pending operations. (See *ESR? for the bit configuration
of this register.) Pending operations are complete when:
•
All previous commands have been executed.
•
Any change in the input level caused by previous commands has been completed.
(Effects of slew rate have been accounted for.)
•
No pending trigger level operations are set for the single electronic load or for any
channel of the multiple electronic load.
*OPC does not prevent processing of subsequent commands but Bit 0 will not be set until
all pending operations are complete.
Command Syntax
*OPC
Parameters
None
Related Commands
*WAI *OPC?