Specifications
Verification and Performance Tests 21
Table 2-5. Performance Test Record for Model Agilent 66104A (60V, 2.5A)
Pre Cal_____________Post Cal______________
Model Agilent 66104A Temperature___________________________Test Performed
By______________________
Serial Number_______________________Humidity________________________Date__________________________
Test Description Specification Actual
CV PROGRAMMING @ 0 volts
Output Accuracy (DVM Reading)
Display and Readback Accuracy
0 ± 27mV
DVM ± 16mV
________
________
CV PROGRAMMING @ full-scale
Output Accuracy (DVM Reading)
Display and Readback Accuracy
60V ± 45mV
DVM ± 28mV
________
________
CV LOAD EFFECT 2mV ________
CV SOURCE EFFECT 2mV ________
CV NOISE (PARD)
Peak to Peak
Rms
15mV
9mV (1)
________
________
TRANSIENT RECOVERY in 1 ms 100mV ________
CC PROGRAMMING @ 0 amps
Output Accuracy (Io)
Display and Readback Accuracy
0 ± 0.6mA
Io ± 0.6mA
________
________
CC PROGRAMMING @ full-scale
Output Accuracy (Io)
Display and Readback Accuracy
2.5A ± 1.4mA
Io ± 1.1mA
________
________
CC LOAD EFFECT 0.1mA ________
CC SOURCE EFFECT 0.1mA ________
CC NOISE (RMS) 1mA ________
(1) 18mV Rms when the ac input is between 87 and 104Vac.