Specifications

20 Verification and Performance Tests
Table 2-4. Performance Test Record for Model Agilent 66103A (35V, 4.5A)
Pre Cal_____________Post Cal______________
Model Agilent 66103A Temperature___________________________Test Performed
By______________________
Serial Number_______________________Humidity________________________Date__________________________
Test Description Specification Actual
CV PROGRAMMING @ 0 volts
Output Accuracy (DVM Reading)
Display and Readback Accuracy
0 ± 13mV
DVM ± 8mV
________
________
CV PROGRAMMING @ full-scale
Output Accuracy (DVM Reading)
Display and Readback Accuracy
35V ± 23.5mV
DVM ± 15mV
________
________
CV LOAD EFFECT 1mV ________
CV SOURCE EFFECT 1mV ________
CV NOISE (PARD)
Peak to Peak
Rms
10mV
5mV
________
________
TRANSIENT RECOVERY in 1 ms 100mV ________
CC PROGRAMMING @ 0 amps
Output Accuracy (Io)
Display and Readback Accuracy
0 ± 2mA
Io ± 2mA
________
________
CC PROGRAMMING @ full-scale
Output Accuracy (Io)
Display and Readback Accuracy
4.5A ± 3.4mA
Io ± 2.9mA
________
________
CC LOAD EFFECT 0.2mA ________
CC SOURCE EFFECT 0.3mA ________
CC NOISE (RMS) 2mA ________