Specifications
18 Verification and Performance Tests
Table 2-2. Performance Test Record for Model Agilent 66101A (8V, 16A)
Pre Cal_____________Post Cal______________
Model Agilent 66101A Temperature___________________________Test Performed
By_______________________
Serial Number_______________________Humidity________________________Date__________________________
Test Description Specification Actual
CV PROGRAMMING @ 0 volts
Output Accuracy (DVM Reading)
Display and Readback Accuracy
0 ± 3mV
DVM ± 2mV
________
________
CV PROGRAMMING @ full-scale
Output Accuracy (DVM Reading)
Display and Readback Accuracy
8V ± 5.4mV
DVM ± 3.6mV
________
________
CV LOAD EFFECT 1mV ________
CV SOURCE EFFECT 0.5mV ________
CV NOISE (PARD)
Peak to Peak
Rms
5mV
2mV
________
________
TRANSIENT RECOVERY in 1 ms 100mV ________
CC PROGRAMMING @ 0 amps
Output Accuracy (Io)
Display and Readback Accuracy
0 ± 6mA
Io ± 6mA
________
________
CC PROGRAMMING @ full-scale
Output Accuracy (Io)
Display and Readback Accuracy
16A ± 10.8mA
Io ± 9.2mA
________
________
CC LOAD EFFECT 0.5mA ________
CC SOURCE EFFECT 0.75mA ________
CC NOISE (RMS) 8mA ________